ICB/PFSA6/RDF/MI/36/16
August 26, 2019
Tender Code: ICB/PFSA6/RDF/MI/36/16
Award Status: Full Winners List
Award Date: 03/14/2018
Tender Award: Download
Unsuccessful Bidders:
Technical Evaluation Result: Download
S.No | Item | Unit | Qty | Bidder | Compliance | Reason Of Non Compliance |
1 | Anesthesia Machine | Each | 4 | |||
Boston Medical FZE | COMPLY | |||||
Suzhou Industrial | NOT COMPLY | No FMHACA Product Certificate | ||||
Quayle Dental | NOT COMPLY | Patient Monitor Not Indicated | ||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
Beijing Aeonmed | NOT COMPLY | Model & Catalogue Not Indicated For Patient Monitor | ||||
Acoma | NOT COMPLY | O2 & N2O Cylinder Not Quoted | ||||
Mindray | COMPLY | |||||
Comen | NOT COMPLY | Two Flow Meters Against Three Flow Meters | ||||
2 | Laundry Machine | Each | 17 | |||
Narula | NOT COMPLY | Catalogue Not Original | ||||
Omedad | COMPLY | |||||
Alliance | COMPLY | |||||
Tolon | NOT COMPLY | Washer Extractor Capacity 40Kg Against 50Kg Per Cycle | ||||
Italclean | NOT COMPLY | Dryer 36kg Against 40kg | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Beardsell | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Suzhou Industrial | NOT COMPLY | Accessories Not Indicated | ||||
3 | Autoclave 75L, Vertical Type | Each | 4 | |||
Appothecaries | NOT COMPLY | List of Spare Part With Their Part Number Is Not Attached | ||||
Beardsell | NOT COMPLY | No Water Softener | ||||
4 | Automatic External Defibrillator | Each | 1 | |||
Mindray | COMPLY | |||||
Beardsell | NOT COMPLY | Catalogue Is Not Supported With Detail Specification | ||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
5 | Capnography | Each | 2 | |||
EDAN | COMPLY | |||||
Mindray | COMPLY | |||||
6 | Cardiotocography Monitor | Each | 2 | |||
Boston Medical FZE | COMPLY | |||||
EDAN | COMPLY | |||||
Appothecaries | NOT COMPLY | Catalogue Is Not Supported With Detail Specification & No Spare Part List With Part Their Number | ||||
7 | CD4 Machine | Each | 2 | |||
BD | COMPLY | |||||
Mindray | NOT COMPLY | Coring Station Not Indicated, Pre Programmed Electronic Pipette Not Quoted. | ||||
8 | Centrifuge | Each | 15 | |||
Quayle Dental | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Fanem | NOT COMPLY | 5000 RPM Against 6000 RPM | ||||
Infinity Medlab | NOT COMPLY | Catalogue Not Original | ||||
Appothecaries | NOT COMPLY | Swing Out Rotor Against Fixed Angle Rotor, List Of Spare Part With Their Part Number Is Not Attached | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Beardsell | NOT COMPLY | Speed 5500 RPM Against 6000RPM | ||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
Suzhou Industrial | COMPLY | |||||
HEMC | NOT COMPLY | Automatic Break and Lid Interlock Not Indicated | ||||
Human | NOT COMPLY | RPM 5000 Against 6000 & No List Of Spare Part With Their Part Number | ||||
9 | Centrifuge, Hematocrite | Each | 10 | |||
HEMC | NOT COMPLY | Automatic Break and Lid Interlock Not Indicated | ||||
Quayle Dental | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Infinity Medlab | NOT COMPLY | Catalogue Not Original | ||||
Nuve | COMPLY | |||||
Appothecaries | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached, Speed RPM 11500 Against 12000 | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Fanem | NOT COMPLY | Adaptor For Smaller Tubes Not Indicated | ||||
Human | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
Beardsell | NOT COMPLY | No FMHACA Product Certificate | ||||
Suzhou Industrial | NOT COMPLY | 200 Capillary Tube Is Not Indicated | ||||
10 | Coagulation Analyzer | Each | 2 | |||
Diagnostic Stago | COMPLY | |||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Suzhou Industrial | NOT COMPLY | 20 Sample Position Against 50 , 40 Sample/Hr Against 260 | ||||
SYSMEX | NOT COMPLY | Throughput 180 Tests Per Hour Against 260 Tests Per Hour | ||||
Human | NOT COMPLY | Sample Position 25 Against 50 | ||||
11 | Flexible Colonoscope | Set | 1 | No bidder | ||
12 | Echocardiography with Cardiac Probe | Each | 1 | |||
Boston Medical FZE | COMPLY | |||||
SonoScape | COMPLY | |||||
Mindray | COMPLY | |||||
13 | Dental Unit | Each | 1 | |||
Quayle Dental | COMPLY | |||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
14 | Color Doppler Ultrasound Machine | Each | 2 | |||
Boston Medical FZE | NOT COMPLY | TEE Probe Not Indicated | ||||
SonoScape | NOT COMPLY | Respiration Trigger Not Indicated | ||||
Mindray | NOT COMPLY | Respiration Trigger Not Indicated | ||||
15 | Fully Digital Radiography Machine (with Out Fluoroscopy) |
Each | 1 | |||
Boston Medical FZE | COMPLY | |||||
Wangdong | NOT COMPLY | Tube Current 25-630mAmp Against 10-8000mAmp | ||||
Suzhou Industrial | NOT COMPLY | Tube Current 10-640mAmp Against 10-800mAmp | ||||
16 | Electrical Centrifuge ,10000 RPM | Each | 2 | |||
Quayle Dental | NOT COMPLY | Rotor 12x15ml Against 12x20ml | ||||
Appothecaries | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Fanem | NOT COMPLY | RPM 5000 Against 10000 | ||||
Beardsell | COMPLY | |||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
Human | NOT COMPLY | RPM 5000 Against 10000 | ||||
17 | Electrical Centrifuge, 24 Tube | Each | 2 | |||
Quayle Dental | NOT COMPLY | 12 Tube Against 24 Tube | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Fanem | NOT COMPLY | No Product Certificate | ||||
Beardsell | NOT COMPLY | 18 Tube Against 24 Tube | ||||
Human | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
18 | Electrical Centrifuge , 36 Tube | Each | 2 | |||
Quayle Dental | NOT COMPLY | 12 Tube Against 36 Tube | ||||
Fanem | NOT COMPLY | No Product Certificate | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Appothecaries | NOT COMPLY | Catalogue Not Attached | ||||
Beardsell | NOT COMPLY | 18 Tube Against 36 Tube | ||||
Human | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
19 | ECG, Six Channel | Each | 1 | |||
SonoScape | COMPLY | |||||
Appothecaries | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached, Paper Size Not A4, Paper Speed Not Indicated In Catalogue | ||||
Boston Medical FZE | COMPLY | |||||
EDAN | COMPLY | |||||
Suzhou Industrial | NOT COMPLY | Accessories Not Indicated | ||||
20 | Electrolyte Analyzer | Each | 2 | |||
Suzhou Industrial | NOT COMPLY | Electrode Mg˖˖, Li˖ Not Indicated, Built In Thermal Printer and Analyze Time Not Indicated | ||||
Human | NOT COMPLY | Electrode Mg˖˖, Li˖ Not Indicated | ||||
21 | Electrosurgical Unit | Each | 15 | |||
HEMC | NOT COMPLY | Cutting Mode Is Not Indicated (Soft, Force,…Bipolar) | ||||
Quayle Dental | NOT COMPLY | Coagulation 150w Against 200w | ||||
Boston Medical FZE | COMPLY | |||||
Acoma | NOT COMPLY | Accessories not Fully Quoted | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Bowa | COMPLY | |||||
Appothecaries | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached, Foot Switch Not Indicated | ||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
Infinity Medlab | NOT COMPLY | Catalogue Not Original | ||||
Medtronic | NOT COMPLY | Mono Foot Switch Against Two Pedal Switch | ||||
22 | Endoscope | Each | 1 | |||
Boston Medical FZE | NOT COMPLY | Laparoscopy Not Indicated | ||||
Medtronic | NOT COMPLY | Colonoscopy Not Indicated | ||||
23 | ESR Analyzer | Each | 2 | |||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Boston Medical FZE | COMPLY | |||||
Suzhou Industrial | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
24 | Glucometer | Each | 17 | |||
Narula | NOT COMPLY | Catalogue Not Original | ||||
25 | Hematology Analyzer 5 Differential | Each | 3 | |||
Mindray | COMPLY | |||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Abbot | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Nihon Kohden | COMPLY | |||||
Human | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Boston Medical FZE | COMPLY | |||||
SYSMEX | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Meril Diagnostic | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached, Installation and After Sales Services Are Not Indicated | ||||
26 | Hematology Analyzer 3 Differential | Each | 5 | |||
Mindray | COMPLY | |||||
Abbot | COMPLY | |||||
Human | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Nihon Kohden | COMPLY | |||||
Boston Medical FZE | COMPLY | |||||
Suzhou Industrial | NOT COMPLY | No Product Certificate | ||||
Meril Diagnostic | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached, Installation and After Sales Services Are Not Indicated | ||||
SYSMEX | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
27 | Hormone Analyzer | Each | 2 | No Bidder | ||
28 | Infant Radiant Warmer | Each | 11 | |||
HEMC | NOT COMPLY | Over Head Exam Light Not Indicated, Preset & Actual Skin, Air Elapsed/Remained Temperature Not Indicated | ||||
Quayle Dental | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Fanem | COMPLY | |||||
Appothecaries | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Ningbo David | COMPLY | |||||
Infinity Medlab | NOT COMPLY | Catalogue Not Original | ||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
Boston Medical FZE | NOT COMPLY | No Integrated Timer | ||||
29 | Infusion Pump | Each | 3 | |||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Boston Medical FZE | COMPLY | |||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
Appothecaries | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached, Battery Life 3 Hr Against 8 Hr | ||||
Suzhou Industrial | NOT COMPLY | Catalogue Not Attached | ||||
30 | Lensometer | Each | 5 | |||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
31 | Medical Refrigerator Large Volume 650L | Each | 8 | |||
Quayle Dental | COMPLY | |||||
ABS International FZE | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
B Medical | COMPLY | |||||
Beardsell | NOT COMPLY | |||||
Boston Medical FZE | COMPLY | |||||
Suzhou Industrial | NOT COMPLY | Power Consumption 443W Against 320W | ||||
32 | Mechanical Ventilator | Each | 10 | |||
Mindray | COMPLY | |||||
Quayle Dental | NOT COMPLY | Catalogue Is Not Supported With Detail Specification | ||||
Boston Medical FZE | COMPLY | |||||
Suzhou Industrial | NOT COMPLY | Catalogue Is Not Supported With Detail Specification | ||||
33 | Patient Monitor | Each | 7 | |||
Suzhou Industrial | NOT COMPLY | Separated CPU Module Is Not Supported | ||||
Mindray | NOT COMPLY | Separated CPU Module Is Not Supported | ||||
Boston Medical FZE | NOT COMPLY | Separated CPU Module Is Not Supported | ||||
Comen | COMPLY | |||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
EDAN | COMPLY | |||||
34 | O.R Light, Mobile | Each | 24 | |||
HEMC | NOT COMPLY | Integrated Battery Is Not Indicated | ||||
Quayle Dental | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Appothecaries | NOT COMPLY | Battery Capacity Not Indicated | ||||
Dr. Mach | COMPLY | |||||
Beardsell | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Boston Medical FZE | COMPLY | |||||
Suzhou Industrial | NOT COMPLY | Built-In Battery Not Indicated | ||||
Mindray | COMPLY | |||||
Infinity Medlab | NOT COMPLY | Catalogue Not Original | ||||
Beijing Aeonmed | COMPLY | |||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
35 | Mobile X-Ray System, Digital | Each | 2 | |||
Boston Medical FZE | COMPLY | |||||
Wangdong | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
SHIMADZU | NOT COMPLY | AEC Not Supported | ||||
Suzhou Industrial | NOT COMPLY | Accessories and Consumables Are Not Indicated | ||||
36 | Microscope, Binocular | Each | 22 | |||
HEMC | NOT COMPLY | Mechanical Stage 120x120mm Against 200x250mm, No Interpupillary Distance and Diopter Adjustment | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Appothecaries | NOT COMPLY | Mechanical Stage and Travelling Distance, Coarse and Fine Focusing, Eye Guard Retractable Not Indicated In Catalogue | ||||
Beijing Teda | NOT COMPLY | Retractable Eye Guard Not Indicated | ||||
Human | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Beardsell | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Infinity Medlab | NOT COMPLY | Catalogue Not Original | ||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
37 | Semen Analyzer | Each | 1 | No Bidder | ||
38 | Generator Set | Each | 1 | |||
Beardsell | COMPLY | Provisionally Accepted | ||||
Suzhou Industrial | COMPLY | |||||
39 | Steam Sterilizer 25 L | Each | 36 | |||
Appothecaries | NOT COMPLY | Power Consumption 1.85kw Against 1.5kw, List Of Spare Part With Their Part Number Is Not Attached | ||||
Nuve | COMPLY | |||||
Beardsell | NOT COMPLY | No FMHACA Product Certificate, No List Of Spare Part With Their Part Number | ||||
Boston Medical FZE | COMPLY | |||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
40 | Contact Tonometer | Each | 5 | No Bidder | ||
41 | Water Bath 8 L with Lid | Each | 7 | |||
Fanem | NOT COMPLY | Quoted Model Not Indicated In Catalogue | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Karl hecht | NOT COMPLY | Bath Volume Not Indicated | ||||
Quayle Dental | COMPLY | |||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
42 | Water Distiller 20L/Hr, with Tank | Each | 7 | |||
Quayle Dental | COMPLY | |||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
43 | Vacuum Extractor Electrical | Each | 105 | |||
HEMC | NOT COMPLY | Modified Bird Cup 60mm Widest, 50cm At Opening and Flexion Point At 6 & 11 Not Indicated | ||||
Appothecaries | NOT COMPLY | Quoted Item Is Not Vacuum Extractor, No Integral Palm Pump | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
Infinity Medlab | NOT COMPLY | Catalogue Not Original | ||||
44 | Blood Bank Refrigerator | Each | 15 | |||
Fanem | NOT COMPLY | Electronic Fuse Auto Disconnect/Connect Not Indicated | ||||
Beardsell | NOT COMPLY | Power Consumption 460W Against 250W | ||||
B Medical | COMPLY | |||||
ABS International FZE | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Nuve | COMPLY | |||||
Quayle Dental | NOT COMPLY | Volume 388L Against 450L | ||||
Boston Medical FZE | COMPLY | |||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Suzhou Industrial | NOT COMPLY | Automatic Voltage Regulator Not Indicated. Ambient Temperature 10-32 Deg C Against 10-43 Deg C | ||||
45 | General Purpose Ultrasound | Each | 17 | |||
Mindray | COMPLY | |||||
SonoScape | COMPLY | |||||
Boston Medical FZE | COMPLY | |||||
EDAN | NOT COMPLY | No Detail Specification, 2 – 6 MHz Against 5 – 7.5 MHz, 6-10MHz Against 5-7.5mhz | ||||
Quayle Dental | NOT COMPLY | Beam former Capacity 3.5 MHz to 7.5 MHz Against 3.0 MHz to 9.0 MHz | ||||
46 | Chemistry Machine | Each | 1 | |||
Human | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Mindray | COMPLY | |||||
Suzhou Industrial | NOT COMPLY | Catalogue is not supported with detail specification | ||||
47 | Surgical Suction Machine, Elec, 2 Bottle | Each | 30 | |||
HEMC | NOT COMPLY | Flow Meter 35L/min Against 60L/min | ||||
Appothecaries | NOT COMPLY | Vacuum Capacity 600mmHg Against 900mmHg, List Of Spare Part With Their Part Number Is Not Attached | ||||
Infinity Medlab | NOT COMPLY | Catalogue Not Original | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
48 | Oxygen Concentrator | Each | 60 | |||
HEMC | NOT COMPLY | Oxygen Concentration: 95% ±3% At 1-3 LPM, 92% ±3% At 4 LPM, 90% ±3% At 5 LPM Is Not Indicated | ||||
Appothecaries | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Longfian | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Infinity Medlab | NOT COMPLY | Catalogue Not Original | ||||
Quayle Dental | NOT COMPLY | Catalogue Not Original | ||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
49 | Pulse Oximeter | Each | 50 | |||
HEMC | NOT COMPLY | High and Low Alarm Setting & Oxygen Concentration Range Is Not Indicated | ||||
Appothecaries | NOT COMPLY | Oxygen Saturation and Pulse Rate Measurement Range Is Not Indicated | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Beijing Teda | COMPLY | |||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
Infinity Medlab | NOT COMPLY | Catalogue Not Original | ||||
Medtronic | COMPLY | |||||
50 | Deep Freezer with Thermometer 350Lit | Each | 4 | |||
Beardsell | NOT COMPLY | Power Consumption 860W Against 500W | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
ABS International FZE | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
B Medical | COMPLY | |||||
Quayle Dental | NOT COMPLY | Ambient Temperature Not Indicated | ||||
Suzhou Industrial | NOT COMPLY | Automatic Voltage Regulator Not Indicated. Ambient Temperature 10-32 Deg C Against 10-43 Deg C | ||||
51 | Refrigerator 250lit | 24 | ||||
Suzhou Industrial | NOT COMPLY | Automatic Voltage Regulator Not Indicated. Ambient Temperature 10-32 Deg C Against 10-43 Deg C | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Fanem | NOT COMPLY | No Auto Electric Disconnect/Connect Electronic Fuse | ||||
B Medical | COMPLY | |||||
ABS International FZE | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Beardsell | COMPLY | |||||
Quayle Dental | NOT COMPLY | Ambient Temperature Not Indicated | ||||
52 | Hygrometer | Each | 12 | No Bidder | ||
53 | Dry Heat Sterilizer 25L | Each | 12 | |||
HEMC | NOT COMPLY | Operating Time Not Indicated | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Beardsell | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached, One Tray Against Three Tray | ||||
Infinity Medlab | NOT COMPLY | Catalogue Not Original | ||||
Appothecaries | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Nuve | COMPLY | |||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
54 | Vital Sign Monitor | Each | 12 | |||
Mindray | COMPLY | |||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Boston Medical FZE | COMPLY | |||||
Appothecaries | NOT COMPLY | NIBP Measurement Is Not Indicated | ||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
Suzhou Industrial | NOT COMPLY | Catalogue Is Not Supported With Detail Specification | ||||
EDAN | NOT COMPLY | Respiration Rate 0 -150 Bpm Against 6 -180 Bpm | ||||
Comen | COMPLY | |||||
55 | Analytical Balance | Each | 9 | |||
Beardsell | COMPLY | |||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
56 | Shaker Rotary Blood Specimen | Each | 10 | |||
Quayle Dental | COMPLY | |||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Human | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Fanem | NOT COMPLY | Quoted Model Not Indicated In Catalogue | ||||
Karl Hecht | NOT COMPLY | 16 X 8 ml Against 20 X 15 ml Test Tube | ||||
57 | Rotary Agglutination Test | Each | 16 | |||
Fanem | NOT COMPLY | Quoted Model Not Indicated In Catalogue | ||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Human | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached & RPM 80 Against 100 | ||||
Quayle Dental | COMPLY | |||||
Narula | NOT COMPLY | Catalogue Not Original | ||||
58 | Vortex, Test Tube | Each | 8 | |||
Quayle Dental | COMPLY | |||||
Karl Hechet | COMPLY | |||||
Avishkar | NOT COMPLY | Catalogue Not Original | ||||
Human | NOT COMPLY | List Of Spare Part With Their Part Number Is Not Attached | ||||
Fanem | NOT COMPLY | Quoted Model Not Indicated In Catalogue | ||||
End of Item Based Result | ||||||
1 | LOT I: Instrument sets(Item 1- 88) | |||||
Quayle Dental | COMPLY | |||||
Avishikar | NOT COMPLY | Sample is not submitted | ||||
Boston Medical FZE | NOT COMPLY | Incomplete lot, sample is not submitted | ||||
Appothecaries | COMPLY | |||||
Trudent | NOT COMPLY | Sample is not submitted | ||||
Deluxe | NOT COMPLY | Incomplete lot, sample is not submitted | ||||
Kay and Company | NOT COMPLY | Sample is not submitted | ||||
2 | LOT II: Laboratory glass wares and instruments (Item 1- 69) | |||||
Deluxe | NOT COMPLY | Incomplete lot, Catalogue not orginal | ||||
ABS international | NOT COMPLY | Incomplete lot | ||||
Quayle Dental | NOT COMPLY | No compliance sheet with specification, orginal manufacturer catalogue is not attached | ||||
Avishikar | NOT COMPLY | Catalogue not orginal | ||||
Boston Medical FZE | NOT COMPLY | Incomplete lot | ||||
Appothecaries | NOT COMPLY | Catalogue not orginal | ||||
Karl Hecht | NOT COMPLY | Incomplete lot | ||||
3 | LOT III: Hospital furnitures (Item 1- 26) | |||||
Quayle Dental | NOT COMPLY | Orginal manufacturer catalogue is not attached for quoted item no 3 & 10 | ||||
Boston Medical FZE | NOT COMPLY | Incomplete lot | ||||
Deluxe | NOT COMPLY | Incomplete lot, Catalogue not orginal | ||||
Appothecaries | NOT COMPLY | Carrying capacity and frame diameter not indicated for item no 5, 8. lifting height, overall size is not indicated for item no 10 | ||||
Avishikar | NOT COMPLY | Catalogue not orginal, copy paste of PFSA specification | ||||
Suzhou Industrial | NOT COMPLY | Incomplete lot | ||||
Beardsell | NOT COMPLY | Different Model in Compliance Sheet and Cattalogue for item 4,11 and 15; Item11 Mortuarry trolley height adjustment 775-1140mm against 400-1725mm | ||||
Linet | NOT COMPLY | Incomplete lot | ||||
Trigenesis | NOT COMPLY | Catalogue not orginal, 200g against 100g for item no 12, 625x475x920mm against 860x560x1040mm for item no 9 | ||||
Kay and Company | NOT COMPLY | Foam density is not indicated, 1830x600x700mm against 1800x800x750mm for Gyn.Ex. coach, and list of spare part for patient bed | ||||
Narula | NOT COMPLY | Catalogue not orginal and model not indicated | ||||
Infinity Medlab | NOT COMPLY | Catalogue not orginal and model not indicated | ||||
HEMC | NOT COMPLY | Weighing adult scale measuring range and graduation not indicated for item no. 12 and baby scale graduation not indicated for item no. 17 | ||||
4 | LOT IV: Diagnostic instruments(Item 1- 21) | |||||
Appothecaries | NOT COMPLY | Orginal manufacturer catalogue is not attaced for item no 4,5,15,17 and 20 | ||||
Narula | NOT COMPLY | Catalogue not orginal and model is not indicated | ||||
Karl Hecht | NOT COMPLY | Incomplete lot | ||||
Trudent | NOT COMPLY | Incomplete lot | ||||
Deluxe | NOT COMPLY | Incomplete lot | ||||
Boston Medical FZE | NOT COMPLY | Incomplete lot | ||||
Avishikar | NOT COMPLY | Catalogue not orginal, Copy paste of PFSA specification. | ||||
Kay and Company | COMPLY | |||||
Infinity Medlab | NOT COMPLY | Catalogue not orginal, model not indicated | ||||
HEMC | NOT COMPLY | Quoted item for item no. 17 is not digital hemoglobinometer | ||||
5 | LOT V: Miscellaneous instruments (Item 1- 24) | |||||
Deluxe | NOT COMPLY | Catalogue not orginal | ||||
Appothecaries | NOT COMPLY | Orginal manufacturing catalogue is not attached for item no 9,10,11 | ||||
Avishikar | NOT COMPLY | Catalogue not orginal |